The micromanipulator and optics
On this page you will learn
- the parts of an ICSI workstation and what each one does
- why modulation contrast optics are used for plastic dishes
- how the manipulators and injectors give you fine control
- the daily start-up checks you do before any case
The workstation as one system
An ICSI workstation, often called the rig, is an inverted microscope fitted with a heated stage, two micromanipulators and two microinjectors, standing on an anti-vibration table. It is usually placed in a quiet corner of the laboratory away from doors and air currents (page 16). Every part must work together: an excellent manipulator is useless if the stage is cold or the optics are out of alignment.
The inverted microscope
An inverted microscope has its objectives below the stage and the light source above. This lets you look up through the bottom of the dish while the pipettes enter the drops from above. You will use a low-power objective to find drops and bring pipettes into position, a middle objective to select sperm and a higher objective, usually around ×40, for immobilisation and injection.
Contrast methods
Oocytes and sperm are nearly transparent, so plain brightfield shows little. Three methods are in use.
- Modulation contrast: gives a shadowed, three-dimensional look and works through plastic dishes. This is the standard for ICSI.
- Differential interference contrast: gives fine detail but needs glass-bottomed dishes, because plastic disturbs the polarised light.
- Polarised light imaging: shows birefringent structures such as the meiotic spindle and the zona layers, and can help to avoid the spindle during injection (page 31).
Each modulation contrast objective has a matching slit in the condenser. If the condenser slit and the objective do not match, or are not aligned, the image looks flat.
The heated stage
The stage carries a heated plate so drops stay at body temperature. The display shows the plate temperature, not the drop. You confirm the drop temperature with a calibrated probe in a dish set up exactly as for a case. The difference between display and drop is recorded and the set point adjusted (page 17).
Micromanipulators
Each side has a coarse control to bring the pipette into the field and a fine joystick for movement in three axes. Fine controls may be hydraulic or electronic. By convention the holding pipette is on the left and the injection pipette on the right; left-handed operators may set up the other way. Each pipette is held in a tool holder at an angle so that its bent tip lies flat to the dish bottom (page 29).
Microinjectors
The microinjectors control suction and pressure in each pipette. The holding side is usually an air-filled syringe, giving gentle steady suction. The injection side needs fine, immediate control, so it is often filled with oil, or uses a precise air system. A screw-driven syringe lets you move fluid a fraction of a turn at a time. Bubbles in an oil-filled line make control spongy and delayed, so they must be removed.
Further tools
Many rigs also carry a laser objective used for assisted hatching and embryo biopsy (pages 56 and 57), and a camera for recording. These are learned later.
At the bench: daily start-up
- Switch on the stage heater early so it is stable before the first case; check drop temperature with the probe and log it.
- Clean the stage plate and condenser surfaces with the approved, embryo-safe method; never use scented wipes or solvents.
- Check the illumination is even and the modulation contrast slit matches the objective in use.
- Centre both joysticks and both coarse controls, so you have full travel in every direction.
- Check both injector lines for bubbles and leaks, and that the screw turns smoothly.
- Confirm the anti-vibration table is floating and nothing touches it.
- Sign the rig start-up log.
Common errors
- Trusting the stage display instead of measuring the drop temperature.
- Starting with a joystick near the end of its travel, then running out of movement mid-injection.
- Leaning on the anti-vibration table or resting equipment on it.
- Using a mismatched contrast slit and blaming the oocytes for a poor image.
Check yourself
- Why is the microscope used for ICSI inverted?
- Why is modulation contrast preferred over differential interference contrast for ICSI dishes?
- What is the difference between the coarse and fine controls of a micromanipulator?
- List the daily start-up checks in order.
Questions on this page
Why is the stage temperature measured in a drop?
The display shows the temperature of the heated plate, not of the medium. Measuring in a drop under oil, set up as for a case, tells you what the oocytes actually experience.
Why is the injection syringe often filled with oil?
Oil does not compress the way air does, so a small turn of the screw gives an immediate, precise movement at the pipette tip. Bubbles spoil this and must be removed.
Will I use the laser on Day 5?
No. The laser is used for assisted hatching and embryo biopsy, taught in Part 5 (pages 56 and 57).
The Genesis ART Textbook is teaching text for trainees and doctors of the Genesis ART Academy. It is not treatment advice and not a laboratory standard operating procedure; every technique is learned under supervision. Page numbers never change.