Pipette alignment and rig set-up
On this page you will learn
- how the holding and injection pipettes are designed
- how to mount, angle and align both pipettes
- how to test suction and pressure before any oocyte is touched
- a repeatable sequence that sets the rig up the same way every time
The two pipettes
The holding pipette has a wide, fire-polished, blunt opening. Gentle suction draws part of the zona against it and holds the oocyte without distorting it. The injection pipette is a fine, bevelled glass needle, often with a small spike at the tip, with an inner diameter just wide enough for one sperm. Both are bought sterile and ready to use, and both have a bend near the tip. Once mounted at the correct angle, this bend makes the last part of each pipette lie flat, parallel to the bottom of the dish. That is the whole aim of alignment.
Why alignment matters
If the injection pipette tip points down, it hits the dish bottom and breaks, or enters the oocyte at an angle and tears it. If it points up, it slides over the zona. If the two pipettes are not in the same focal plane, you hold the oocyte at one level and inject at another, and the oolemma is pushed rather than pierced. A rig that is set up well makes injection gentle; one that is set up poorly makes good technique impossible.
At the bench: set-up sequence
1. Prepare
Complete the daily start-up (page 28). Take the ICSI dish prepared as on page 19, with its sperm drop, PVP drop and oocyte drops of buffered medium under oil, and place it on the heated stage to warm before you need it. Check its label against the worksheet.
2. Mount the pipettes
- Open each pipette pack only when you are ready, and never touch the glass near the tip.
- Insert each pipette into its tool holder and tighten firmly but not so hard that the glass cracks.
- Set the tool-holder angle so that the bent tip is horizontal. Look from the side with your eye at stage level: the tip should look flat.
- Also check the pipette in the horizontal plane: the tip should point straight along the axis of movement, not off to one side.
3. Find the tips
- With the low-power objective, bring each tip into the light path above the dish using the coarse controls. Find the shadow first, then focus up to the tip.
- Lower each tip into an empty area of a drop. Then, with care, lower towards the dish bottom until the tip is just above it.
- Bring both tips to the centre of the field at the same focal plane.
- Change to the injection objective and correct the focus and position again. Both tips should be sharp at the same time.
4. Fill and test the lines
- Holding pipette: apply gentle suction in a medium drop and watch fluid enter smoothly; release and watch it stop. A slow drift means air leak or uneven pressure; correct it before going further.
- Injection pipette: in the PVP drop, draw in a little PVP and push it out again with very small turns. The meniscus should respond at once and stop where you leave it. If it keeps moving, there is a bubble or leak.
- Rotate the injection pipette slightly if needed so the bevel faces the direction your protocol prefers.
5. Final check
Move both pipettes between drops and back without hitting the dish. Lift them out of the drops before you move the stage to a different area. Record the pipette lot numbers on the worksheet for traceability.
Common errors
- Moving the stage while the tips are still low in a drop, so they crash into a drop wall or the dish bottom.
- Aligning only at low power and finding the tips at different heights at ×40.
- Ignoring a drifting meniscus and starting the case anyway.
- Touching the pipette tip with gloves or the packet, which blunts or contaminates it.
- Tightening the holder so hard that the pipette cracks inside it.
- Forgetting to record pipette lot numbers.
How to practise
Set the rig up from scratch, take everything down, and set it up again, timing yourself. Your trainer will check the angle, the focal plane and the flow control. On Day 5 you will repeat this until it is routine, before working with sperm (page 30).
Check yourself
- What is the purpose of the bend near the tip of each pipette?
- Describe how you would check the tip is horizontal.
- Why do you re-check alignment at the injection objective?
- The injection meniscus keeps moving after you stop turning the screw. What does that tell you?
Questions on this page
Can I re-use an injection pipette for the next patient?
No. Injection and holding pipettes are single-use and are changed for each patient, so that gametes from different patients cannot be mixed.
How long should rig set-up take?
A trainee is slow at first. With practice, set-up becomes a short, calm routine that is complete before the oocytes leave the incubator. Speed comes from repetition, not from rushing.
Why record pipette lot numbers?
If a batch of consumables is later found to be faulty, the laboratory can trace every case that used it. See page 55 on traceability.
The Genesis ART Textbook is teaching text for trainees and doctors of the Genesis ART Academy. It is not treatment advice and not a laboratory standard operating procedure; every technique is learned under supervision. Page numbers never change.