Incubators: temperature, gas and pH

On this page you will learn

What an incubator does

An incubator copies a few key conditions of the female reproductive tract: body temperature, a steady pH and, in most laboratories, a reduced oxygen level. Embryos cannot adjust to sudden changes in these conditions. A few minutes on a cold surface, or a dish drifting in pH while the door is open, is stress the embryo has to recover from.

Temperature

The culture target is close to body temperature, usually set at about 37 °C. What matters is the temperature inside the medium drop, not the number on the display. The two can differ, so the laboratory measures temperature inside a dish of medium with a calibrated probe and adjusts the set point accordingly. Heated stages and work surfaces are checked the same way.

Carbon dioxide and pH

Most culture media are buffered with bicarbonate. Bicarbonate reacts with carbon dioxide dissolved in the medium, and the balance between them sets the pH. More carbon dioxide in the incubator lowers the pH; less raises it. So the carbon dioxide set point is not a fixed number for every laboratory: it is chosen to give the pH that the medium manufacturer specifies. It also depends on altitude, because the partial pressure of a gas changes with atmospheric pressure.

Handling media used outside the incubator are different: they use a chemical buffer (such as HEPES or MOPS) that holds the pH in room air. These media must not be placed in a carbon dioxide incubator for long periods, because their pH will fall.

Oxygen

Room air holds about a fifth oxygen. The oviduct and uterus hold much less. Many laboratories culture embryos at reduced oxygen, roughly a quarter of the level in room air, by adding nitrogen. This is thought to reduce oxidative stress. Reduced oxygen incubators need a nitrogen supply and an oxygen sensor, both checked as part of daily QC (page 20).

Types of incubator

Humidity

Some incubators are humidified with a water tray; others are dry. In a dry incubator, the oil overlay (page 18) is what stops the medium from evaporating, so its volume and quality matter even more. A humidified incubator needs its water changed and the chamber cleaned at set intervals to avoid microbial growth.

At the bench

  1. Before opening any incubator, plan what you need so the door is open once and briefly.
  2. Check the display and the alarm status before you open the door. Do not open a chamber that is recovering unless it is urgent.
  3. Move dishes on a warmed carrier or heated block, never on a cold bench.
  4. Return dishes promptly and confirm the chamber and position on your patient record.
  5. Each patient's dishes are in their own labelled space; never place two patients' dishes where they can be confused.
  6. If an alarm sounds, follow your laboratory's alarm procedure: check, record and, if needed, move dishes to a back-up incubator.

The incubators also need regular cleaning, gas cylinder checks, and filter changes. These are logged so that any change in embryo development can be traced back to the equipment if needed.

Common errors

Check yourself

  1. Why does adding carbon dioxide to the incubator lower the pH of the culture medium?
  2. What is the difference between a bicarbonate-buffered culture medium and a HEPES-buffered handling medium?
  3. Give one advantage of a benchtop incubator over a large-box incubator.
  4. In a dry incubator, what protects the culture drops from evaporation?

Contents and reading plan

Questions on this page

Is the carbon dioxide setting the same in every laboratory?

No. It is set to give the pH the medium manufacturer specifies, and that can vary with the medium and with altitude.

Why use reduced oxygen?

It is closer to the conditions in the female reproductive tract and is thought to reduce oxidative stress on embryos. Your laboratory protocol states the set point.

The Genesis ART Textbook is teaching text for trainees and doctors of the Genesis ART Academy. It is not treatment advice and not a laboratory standard operating procedure; every technique is learned under supervision. Page numbers never change.