Andrology QC and troubleshooting

On this page you will learn

Why quality control matters

A semen analysis is used to make decisions about treatment. If two technicians read the same sample and report very different results, the report is not useful. Quality control is the routine work that shows your results are consistent from day to day and close to the true value. The WHO laboratory manual for human semen (6th edition, 2021) gives detailed guidance; your laboratory protocol turns it into daily practice.

Internal quality control

Internal quality control is what you do inside your own laboratory.

External quality assessment

In an external scheme, a provider sends the same material to many laboratories, and each reports its results. You then see how your results compare with the scheme's target values. It is the only way to find a bias that your whole team shares. Treat a poor external result as a learning event: investigate, correct and record.

Equipment and materials

At the bench: troubleshooting

When results go wrong, work from the simple to the complex. Change one thing at a time and write down what you find.

  1. Poor motility after preparation: check the temperature of the warm block and medium, the time from collection to preparation, the centrifuge speed and duration, and the medium lot. Did the collection container come from an approved stock? Some containers and lubricants harm sperm.
  2. Low recovery: check the initial sample assessment, the gradient volumes, how the pellet was collected and the resuspension volume. Was swim-up chosen for a sample better suited to a gradient (page 13)?
  3. Discordant replicate counts: check mixing, chamber loading, chamber depth and the dilution calculation.
  4. Contamination: check aseptic technique, the expiry of media, and the cleaning log for the hood and centrifuge.
  5. A pattern across many samples: suspect something shared, such as a medium lot, a piece of equipment or a change in room conditions. Report it to the person responsible for the laboratory the same day.

Write it down

Every non-conformance (anything that did not go to protocol) is recorded with the cause found and the action taken. These records are part of the quality system the ART (Regulation) Act 2021 and ICMR national ART guidelines expect, and they are how the laboratory improves. The IVF laboratory uses the same logic on pages 17 to 20.

Common errors

Check yourself

  1. What is the difference between internal quality control and external quality assessment?
  2. Two replicate counts disagree by more than the allowed limit. What do you do?
  3. Motility falls sharply after preparation in three samples on the same day. Where do you look first?
  4. Why must the medium lot number be recorded for every preparation?

Contents and reading plan

Questions on this page

How often should observer comparisons be done?

At the interval your laboratory protocol sets. Regular comparison matters more than any single exercise.

What is a non-conformance?

Any event that did not go according to protocol. It is recorded with the cause found and the action taken, so the laboratory can learn from it.