Witnessing, traceability and cryo audits

On this page you will learn

Why witnessing exists

A mix-up of gametes or embryos is among the most serious errors a laboratory can make, and it cannot be undone. It is almost never caused by a lack of skill. It happens when a busy, tired person picks up the wrong tube, reads a label too quickly or puts a dish back in the wrong place. Witnessing is the system that catches those slips before they matter. You met it on page 3 and used it on every day of the course; this page brings the rules together.

Critical points

A critical point is any step where gametes or embryos are moved from one labelled container to another, or where samples from two people meet. These include:

What a witness does

A witness is a second trained person who checks independently. That means the witness reads the labels and the record for themselves; they do not simply listen and agree. Both identifiers are checked on every container, at the moment of the step, not before or after. The witness then signs or confirms in the record, with the time. If the witness is interrupted or called away, the step waits. Electronic witnessing systems (page 60) can take on part of this work, but the principle is the same.

Traceability

Traceability means that for every sample you can answer three questions from the records alone: where is it now, what has been done to it, and who did each step. It also covers everything that touched the sample: culture medium and solution lot numbers, carrier device batches, the incubator and the tank. If a solution lot is later found faulty, traceability lets the laboratory find every sample that met it. The ART (Regulation) Act 2021 and the ICMR national ART guidelines expect records that can show this, and the clinic must keep them for the period set by law.

At the bench: the cryo-inventory audit

On Day 11 you carry out a cryo-inventory audit on a practice tank. A real audit follows the same steps.

  1. Plan the audit with the senior embryologist: which tanks, which canisters, and who will act as checker and recorder.
  2. Print or open the storage register for the section to be audited.
  3. Lift one canister at a time, keeping it in the vapour, and read each device label with your partner. Do not remove devices unless the protocol requires it.
  4. Tick each device against the register: right patient, right address, right sample type, right number of devices.
  5. Note any device that is in the tank but not in the register, any record without a device, any unreadable label and any device in the wrong position.
  6. Check the consent and storage dates for each patient and list any that are due for renewal or past their storage period.
  7. Sign the audit record and report findings the same day.

Audit a section at a time, so that no canister stays out for long. Full audits are done on a schedule set by the laboratory, with smaller spot checks between them.

When something does not match

Do not correct the register or move a device on your own. A discrepancy is reported to the senior embryologist, recorded and investigated: the records of every step for that sample are traced back until the cause is found. Unreadable labels are dealt with by a documented re-identification process, never by guesswork. The aim of a good audit culture is that people report problems early, so the record must never be quietly tidied.

Your logbook

Every witnessed step you perform during the course goes in your logbook (page 61) with the name of your witness. Being a reliable witness is a skill in its own right, and it is assessed.

Common errors

Check yourself

  1. List five critical points in the laboratory that must be witnessed.
  2. What does "independent" checking mean for a witness?
  3. What three questions should traceability let you answer for any sample?
  4. You find a straw in the tank that is not in the register. What do you do?

Contents and reading plan

Questions on this page

Can the same person who performed a step also witness it?

No. The witness must be a second trained person who checks the labels and record independently.

How often should a cryo audit be done?

On a schedule set by the laboratory and its quality system, with spot checks between full audits. The schedule should be written down and kept.

What if a label has become unreadable?

Report it. The sample is re-identified through a documented process using the storage records, with senior oversight. It is never guessed.

The Genesis ART Textbook is teaching text for trainees and doctors of the Genesis ART Academy. It is not treatment advice and not a laboratory standard operating procedure; every technique is learned under supervision. Page numbers never change.