Page 20 of 72Genesis ART Textbook
Part 2: The IVF laboratory, oocytes and ICSI · Day 3
Page 20 of 72

Daily QC and lab KPIs (Vienna consensus)

On this page you will learn

Three ideas, one aim

All three serve one aim: every patient's gametes and embryos get the same reliable care, every day.

The daily checks

Each morning, before any patient work, the laboratory records a set of readings. Typical checks include:

Every reading is written in the log with the time and your initials. A reading out of range is acted on before any work in that equipment, and the action is recorded.

The Vienna consensus

In 2017, ESHRE and Alpha (Scientists in Reproductive Medicine) published the Vienna consensus on ART laboratory performance indicators. It sets out a group of indicators that cover the main steps of laboratory work, for example: the proportion of oocytes damaged by ICSI, normal fertilisation after ICSI and after conventional IVF, cleavage on Day 2, blastocyst development, survival of embryos after warming, and implantation. For each indicator it suggests a "competency" value, which every laboratory should reach, and a "benchmark" level, which is an aspirational target.

It also separates reference indicators (such as the proportion of mature oocytes at pick-up, which depends on the clinic as well as the laboratory), performance indicators (which mainly reflect laboratory work) and outcome indicators. This helps you work out where a problem truly lies.

Using KPIs well

At the bench

  1. Arrive in time to complete the QC checks before the first case.
  2. Measure; do not copy yesterday's number. Write the actual reading.
  3. If a reading is out of range, repeat it once with care. If still out, inform the senior embryologist, move dishes to a back-up if needed and record what you did.
  4. Enter outcomes for each case (fertilisation, development, cryo) on the day, so KPI data are complete.
  5. Take part in the monthly KPI review and ask how your own work compares with the laboratory's competency values.

Quality systems are also expected under the ART (Regulation) Act 2021 and the ICMR national ART guidelines, so the records you keep are part of the laboratory's legal duty, not only good habit.

Common errors

Check yourself

  1. What is the difference between quality control and quality assurance?
  2. Name five readings taken before patient work begins each morning.
  3. What is the difference between a competency value and a benchmark value in the Vienna consensus?
  4. Why is the proportion of mature oocytes a reference indicator and not a pure laboratory performance indicator?

Contents and reading plan

Questions on this page

Are KPIs figures to show patients?

No. KPIs measure how well each step of laboratory work is done, so problems can be found and fixed. They are internal quality tools.

What if a QC reading is out of range?

Repeat it carefully, inform the senior embryologist, protect the dishes and record the reading, the cause and the action taken.

The Genesis ART Textbook is teaching text for trainees and doctors of the Genesis ART Academy. It is not treatment advice and not a laboratory standard operating procedure; every technique is learned under supervision. Page numbers never change.