IVF centre in Goa
Consultations in Panjim · Margao · Mapusa · Vasco
Call
Page 9 of 72Genesis ART Textbook
Part 1: Andrology, semen preparation and IUI · Day 1
Page 9 of 72

Reporting a semen analysis

On this page you will learn

The purpose of the report

A semen analysis report is read by a doctor who was not at your bench. It must show exactly what you measured, how and when, so that the doctor can trust it and compare it with later tests. Clear facts are more useful than labels.

What the report contains

Identification and collection

Macroscopic results (page 6)

Liquefaction time, appearance, viscosity, volume and pH.

Microscopic results (pages 7-8)

Sperm concentration and total sperm number; motility in the four categories, with total and progressive motility; vitality, if done; normal forms and the main defects; round cells and leucocytes; agglutination or aggregation, if seen. Give every value with its unit, and state the method where your laboratory has more than one (for example manual or CASA).

Reference limits

The WHO laboratory manual for human semen (6th edition, 2021) gives lower reference limits drawn from men whose partners conceived within a year of trying. Each limit is the value that only the lowest twentieth of those men fell below (the fifth centile). The main limits in the 6th edition are:

MeasureLower reference limit
Semen volume1.4 millilitres
Sperm concentration16 million per millilitre
Total sperm number39 million per ejaculate
Total motility42 in every 100 spermatozoa
Progressive motility30 in every 100 spermatozoa
Vitality (live)54 in every 100 spermatozoa
Normal forms4 in every 100 spermatozoa

Read these limits with care. A result below a limit does not mean a man cannot father a child, and a result above every limit does not prove that he can. The 6th edition itself warns against using the limits as a fertile or infertile cut-off. Always check the figures against the manual on your bench, and print them on the report only in the form your laboratory has approved.

Descriptive terms

These standard terms are a shorthand. Use them only alongside the actual numbers.

Terms can be combined, for example oligoasthenoteratozoospermia, but long labels can sound alarming to a couple. The report is for the doctor, who explains it.

Why one test is not enough

Semen quality varies a great deal from week to week in the same man. A single abnormal result is usually repeated before conclusions are drawn. Your report should make repeat testing easy to compare: same method, same units, same layout.

At the bench: checking and releasing a report

  1. Enter results directly from the worksheet; do not rely on memory.
  2. Check that the calculations are right: total sperm number equals concentration times volume.
  3. Check internal consistency: live sperm not fewer than motile sperm; motility categories adding up to the whole; a clear sample with a low count.
  4. Check identity on the report against the worksheet and the sample label.
  5. A second trained person reviews the report before release.
  6. Release the report only through the approved route, to the requesting doctor. Semen results are confidential.

Common errors

Check yourself

  1. List five items that must appear in the identification section of a report.
  2. What does a fifth-centile lower reference limit mean?
  3. Define cryptozoospermia and azoospermia, and explain why the difference matters.
  4. Name two internal consistency checks you make before a report is released.

Contents and reading plan

Questions on this page

Is a man with results below the reference limits infertile?

Not necessarily. The limits describe fertile men, not a dividing line. Many men below a limit father children, and the doctor interprets results with the whole history.

Why repeat a semen analysis?

Semen quality varies from week to week in the same man, so one abnormal result is usually repeated before conclusions are drawn.

The Genesis ART Textbook is teaching text for trainees and doctors of the Genesis ART Academy. It is not treatment advice and not a laboratory standard operating procedure; every technique is learned under supervision. Page numbers never change.