Single density gradient II: the steps, one by one
On this page you will learn
- each step of the single-layer gradient, in order
- the typical volumes, forces and times, and why your laboratory sets its own
- how to remove the upper layers without losing the pellet
- where the witness checks come
Before step 1
Your kit is warm, the quality-control log is filled and the sample has liquefied and been checked (page 26). Every number below is typical; follow the Heartbeats lab SOP (standard operating procedure). The World Health Organization (WHO) manual asks each laboratory to set and test the force and time that give a pellet it can handle, and to test any change before it is used for patients.
Source: 2021, World Health Organization - WHO laboratory manual for the examination and processing of human semen, sixth edition, section 5.1.1.1
The steps
- Label two sterile conical tubes on their bodies, gradient and wash, with your witness.
- Pipette the warmed colloid into the gradient tube. Typical volume: about 0.5 to 1.5 millilitres.
- Mix the semen gently. Hold the tube at a slight angle and let the semen run slowly down the wall onto the colloid. A clear line should stay between them. Typical: no more semen than about the colloid volume; use a second tube for a large sample.
- Cap the tube and place it in a sealed bucket, with a balance tube of equal weight opposite.
- Centrifuge. Typical: 300 to 400 g for 15 to 30 minutes, at the setting in the SOP.
- Wait for the rotor to stop before you open the lid. Check that a pellet is visible.
- With a fresh pipette, take off the seminal plasma and the upper part of the colloid from the top down. Leave the pellet and a small volume of fluid above it.
- With another fresh pipette, take up the pellet and move it into the wash tube holding warm medium. Typical: about 2 to 5 millilitres. Your witness checks both labels.
- Mix gently, then spin again. Typical: about 200 g for 4 to 10 minutes. Some laboratories wash twice.
- Remove the supernatant carefully, without touching the pellet.
- Resuspend the pellet in the final volume for intrauterine insemination (IUI). Typical: about 0.2 to 0.5 millilitre, or what the catheter and the SOP set.
- Take a small drop for the post-wash check (page 28). Label the final tube with your witness and record every time on the worksheet.
Source: 2010, World Health Organization - WHO laboratory manual for the examination and processing of human semen, fifth edition, section 5.5 (replaced by the 2021 edition; cited for its worked two-layer values)
Source: Jun 2022, Hoa NT et al. (Acta Informatica Medica 30(2):100-104) - The effectiveness of sperm preparation using density mini-gradient and single-layer centrifugation for oligospermia samples
Why each step matters
- Slow layering keeps the line between semen and colloid. Mixing them destroys the separation.
- The force and time are a balance: too little and motile sperm stay in the layer; too much and dead cells and debris are pushed down with them.
- A fresh pipette for the pellet stops debris from the upper layers being carried down.
- The wash spin removes the colloid; skipping it puts colloid into the uterus.
- The final volume must be small enough for the catheter and the uterus, and large enough to load without loss.
Source: 2021, World Health Organization - WHO laboratory manual for the examination and processing of human semen, sixth edition, section 5.5
Force, not speed
Values are written as relative centrifugal force (g), not revolutions per minute (rpm). The same rpm gives a different force in a rotor of a different size. Every centrifuge carries its own chart, so read the rpm for the force in the SOP from the chart beside that machine.
Source: 2021, World Health Organization - WHO laboratory manual for the examination and processing of human semen, sixth edition, section 5.3
At the bench
On Day DI-2 the faculty performs the whole preparation once while you watch and call out each step. You then do preparation 1 under direct supervision, logged as Assisted. Preparations 2 and 3 follow on Day DI-3.
Common mistakes
- Pouring semen onto the colloid instead of letting it run down the wall.
- Spinning without a balance tube.
- Opening the lid while the rotor still turns.
- Using one pipette for both the upper layers and the pellet.
- Resuspending in more volume than the IUI needs.
Key points
- Order: layer, spin, take off the top, move the pellet, wash, resuspend, check.
- Numbers: all values are typical; your SOP rules.
- Pipettes: a fresh one for the pellet.
- Witness: at every tube change and at the final label.
Check yourself
Answer in your head first, then tap a question to see the short answer.
1. Why must semen be layered slowly onto the colloid?
To keep a clear line between them; mixing ruins the separation.
2. What can go wrong if the gradient spin is too long or too strong?
Dead cells and debris are pushed into the pellet with the motile sperm.
3. Why is a fresh pipette used to take up the pellet?
So that debris and cells from the upper layers are not carried down into the pellet.
4. Why is the pellet washed before IUI?
To remove the colloid, which should not go into the uterus.
5. Why are forces written in g rather than rpm?
Because the same rpm gives different forces in rotors of different sizes.
6. From page 26: what should you do if a sample has not liquefied after 30 minutes?
Tell the embryologist; do not start the gradient.
Training note. Training content for registered medical practitioners. Trainees observe all patient care; they do not treat patients during the programme.
Questions on this page
Do doctors perform this preparation on patient samples?
During the programme, doctor trainees prepare training samples under supervision; the IUI on a patient is observed.
Why are the numbers called typical?
Each laboratory tests and fixes its own volumes, forces and times for its own centrifuge and products, as the WHO manual asks.
The Genesis Doctors' Textbook is teaching text for registered medical practitioners on the doctor programmes of the Genesis ART Training Academy. It is not treatment advice and gives no medicine doses; trainees observe all patient care, ultrasound scans are watched only (PCPNDT Act), and bench skills are practised under supervision. Page numbers never change.