Single density gradient III: post-wash check, total motile count, fit for use, errors
On this page you will learn
- what is measured after the wash
- how to calculate the total motile count
- who decides that a sample is fit for use
- how the intrauterine insemination (IUI) sample is labelled and handed over
- the common errors and how to rescue a poor recovery
The post-wash check
Mix the final suspension gently and load a counting chamber. Record the sperm concentration, the percentage of progressively motile sperm and the final volume. Note anything unusual: clumping, many round cells or visible debris. Write the time. The same drop tells you whether the preparation worked and gives the numbers the clinician will read.
Source: 2021, World Health Organization - WHO laboratory manual for the examination and processing of human semen, sixth edition, section 5.4.2
The total motile count
The total progressively motile count tells you how many forward-moving sperm will go into the uterus. Multiply three numbers: the final volume in millilitres, the concentration in millions per millilitre and the fraction that is progressively motile (write it as a decimal).
Worked example. Final volume half a millilitre; concentration 20 million per millilitre; progressive fraction 0.8 (eight sperm in ten move forward). Total progressively motile count = 0.5 × 20 × 0.8 = 8 million.
Do the same sum on the raw semen and compare. The ratio of the two shows how much your preparation recovered, and a falling ratio over several preparations is an early warning about your technique or your equipment.
Is there a lower limit for IUI?
Many units use a lower limit for the post-wash count, but the studies behind such limits disagree. A systematic review found that the cut-offs reported for the inseminated motile count ranged widely, from under one million to about five million, because studies used different methods, patient groups and treatment plans. Each unit therefore writes its own rule. When a count falls below it, the next step is a conversation, not a silent insemination.
Source: 2014, Ombelet W et al. (Reproductive BioMedicine Online 28(3):300-309) - Semen quality and prediction of IUI success in male subfertility: a systematic review
Fit for use: who decides
The embryologist reports the result and says whether the sample meets the unit's criteria. The treating clinician decides, with the couple, whether to go ahead, to repeat in another cycle or to consider in vitro fertilisation (IVF) or intracytoplasmic sperm injection (ICSI). Neither decides alone, and the reasoning is written in the cycle record.
Labelling and hand-over
- Label the final tube with two identifiers of the couple and the unique code, with your witness.
- Write the post-wash report: volume, concentration, progressive motility, total motile count and time.
- At hand-over, read the identifiers aloud with the clinician; the woman states her own full name and date of birth.
- Keep the sample at the temperature in the standard operating procedure (SOP), away from heat and light, and use it within the time window the SOP sets.
Source: Dec 2015, ESHRE Guideline Group on good practice in IVF labs - Revised guidelines for good practice in IVF laboratories (2015)
Errors and rescue
| Error | What you see | Prevention or rescue |
|---|---|---|
| Pellet lost with the supernatant | Very few sperm in the final drop | Keep every removed fluid in a labelled tube until the final check; re-spin it if needed |
| Mixed layers | Debris and round cells after washing | Layer slowly; if mixed, start again in a fresh tube |
| Spin too long or too strong | Dense pellet with debris and dead sperm | Check the revolutions per minute (rpm) chart and the timer before each run |
| Cold medium | Sluggish motility at the check | Warm media first; record the warm block reading |
| Wrong or missing label | Labels on tubes do not match | Stop. The sample is not used until identity is proved |
When numbers are very low, the laboratory may change the force or time to recover more sperm, but always in ways it has tested before. A poor recovery from a good raw sample points to the technique; a poor recovery from a poor raw sample points to the sample.
Source: 2021, World Health Organization - WHO laboratory manual for the examination and processing of human semen, sixth edition, section 5.1.1.1
At the bench
You perform preparations 2 to 5 yourself, on training samples. Each one ends with your own post-wash check, total motile count and hand-over read-back, and each row in your logbook is signed by you and the faculty. The target is Independent at the last preparation.
Common mistakes
- Forgetting the motility fraction in the count sum.
- Discarding the supernatant before the final check.
- Deciding "fit for use" alone, without the clinician.
- A silent hand-over with no read-back.
Key points
- Check: concentration, progressive motility, volume, time.
- Count: volume × concentration × progressive fraction.
- Decision: embryologist reports, clinician and couple decide.
- Identity: a mismatch stops the IUI.
Check yourself
Answer in your head first, then tap a question to see the short answer.
1. Calculate: 0.4 millilitre, 15 million per millilitre, progressive fraction 0.7. What is the total progressively motile count?
0.4 × 15 × 0.7 = 4.2 million.
2. Why compare the pre-wash and post-wash counts?
The ratio shows how much the preparation recovered and gives early warning of a technique or equipment problem.
3. Is there one agreed lower limit for IUI?
No. Studies disagree; each unit sets its own rule and discusses low counts with the couple.
4. Who decides that a sample is used?
The clinician with the couple, on the embryologist's report and the unit's criteria.
5. Why keep the removed fluids until the final check?
If the pellet was lost, the sperm can still be recovered by spinning those fluids again.
6. From page 27: which step removes the colloid?
The wash spin, after the pellet has been moved into fresh medium.
Training note. Training content for registered medical practitioners. Trainees observe all patient care; they do not treat patients during the programme.
Questions on this page
What is the total motile count?
The number of progressively motile sperm in the prepared sample: final volume times concentration times the progressive fraction.
Who decides whether a prepared sample is good enough for IUI?
The embryologist reports against the unit's criteria; the treating clinician decides with the couple.
The Genesis Doctors' Textbook is teaching text for registered medical practitioners on the doctor programmes of the Genesis ART Training Academy. It is not treatment advice and gives no medicine doses; trainees observe all patient care, ultrasound scans are watched only (PCPNDT Act), and bench skills are practised under supervision. Page numbers never change.