Transfer catheters and loading

On this page you will learn

The catheter

An embryo transfer catheter is a sterile, single-use device made of soft, smooth plastic. Most sets have two parts. The outer sheath (or guide) is a little stiffer and can be passed through the cervix first if the canal is tight or curved. The inner catheter is very soft and carries the embryo; it has a fine, rounded, open tip. Many inner catheters have a marked tip that shows clearly on ultrasound, so the clinician can see where the embryo is placed. Every catheter must be embryo-tested, within its expiry date, and opened only once the transfer is about to happen.

The catheter is connected to a small sterile syringe, which lets you draw up and expel very small volumes under fine control. Air in the syringe barrel acts as a cushion; push and pull slowly.

Before loading

The loading sequence

Laboratories use slightly different loading patterns; learn your laboratory's written method and follow it every time. A common pattern loads the embryo in a short column of medium with a small air bubble on either side, which helps the clinician see the column on ultrasound and keeps the embryo in place:

  1. Attach the syringe to the inner catheter and flush it with transfer medium to wet the lumen and remove any residue. Expel the medium.
  2. Draw up a column of medium, then a small air bubble.
  3. Under the microscope, bring the catheter tip close to the embryo and draw it up in a small volume of medium. Watch it enter the tip.
  4. Draw a second small air bubble, then a short column of medium.
  5. Check the catheter under the microscope: the embryo is within the medium column, and the columns are where you expect them. The total volume is small, in the tens of microlitres.
  6. Hand the catheter to the clinician holding it level and steady, tip protected, and say clearly that it is loaded and with which embryo.

From loading to hand-over should take well under a minute. Do not rush, but do not pause. If anything is wrong, unload into the dish and start again with a new catheter.

After transfer: the check

  1. When the catheter comes back, flush it under the microscope into a clean dish of medium.
  2. Search the dish carefully for the embryo, and look at the tip for mucus or blood.
  3. If the embryo is not there, tell the clinician "clear". If it is there, say so at once and calmly: the clinician will decide how to reload.
  4. Record the result and the time on the transfer sheet.

At the bench

In the Genesis Day 8 practical you load practice catheters with training beads or library material, again and again, until the sequence is smooth. You are assessed on clean columns, a visible embryo position and steady hand-over. You then watch a live transfer, observed only, from the laboratory side.

Common errors

Check yourself

  1. What are the two parts of a typical transfer catheter, and what does each do?
  2. Why do many laboratories place small air bubbles either side of the embryo column?
  3. What do you do if the embryo is still in the returned catheter?
  4. Name three checks you make before you start loading.

Contents and reading plan

Questions on this page

Why is the inner catheter so soft?

A soft tip passes through the cervix with less trauma and is less likely to touch the endometrium hard, which can cause bleeding or contractions.

How long should loading take?

From drawing up the embryo to handing the catheter to the clinician should be well under a minute, and loading begins only when the clinician is ready.

Is the returned catheter always checked?

Yes. It is flushed into a dish and searched under the microscope every time, so that a retained embryo is never missed.

The Genesis ART Textbook is teaching text for trainees and doctors of the Genesis ART Academy. It is not treatment advice and not a laboratory standard operating procedure; every technique is learned under supervision. Page numbers never change.